Rigol DSG3000 Series Especificações Página 32

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 31
RIGOL
DS6000 Users Guide
1-10
Function Inspection
1. Press SAVE Factory to restore the oscilloscope to its default configuration.
2. Use the probe to connect the input terminal of channel 1 (CH1) of the
oscilloscope and the “Compensation Signal Output Terminal” of the probe.
Figure 1-8 To Use the Compensation Signal
3. Connect the earth alligator clip of the probe to theGround Terminal” under the
probe compensation signal output terminal.
4. Press the Auto key.
5. Observe the waveform on the display. In normal condition, the display should be
a square waveform as shown in the figure below:
Figure 1-9 Square Waveform Signal
6. Use the same method to test the other channels. If the square waveforms
actually shown do not match that in the figure above, please perform Probe
Compensation” in the next section.
Compensation Signal
Output Terminal
Ground Terminal
Vista de página 31
1 2 ... 27 28 29 30 31 32 33 34 35 36 37 ... 261 262

Comentários a estes Manuais

Sem comentários