RIGOL Chapter 1 Specifications
Service Guide for DS1000E, DS1000D Series
1-4
Trigger Offset
Normal mode: pre-trigger (storage depth/ 2*sampling
rate), delayed trigger 1s
Slow Scan mode: pre-trigger 6div, delayed trigger 6div
Trigger Holdoff range
500ns~1.5s
Set Level to 50%
(Typical)
Input signal frequency ≥50Hz
Edge Trigger
Edge trigger slope
Rising, Falling, Rising + Falling
Pulse Trigger
Trigger condition
(, , =) Positive pulse, (, , =) negative pulse
Pulse Width range
20ns ~10s
Video Trigger
Video standard &
line frequency
Support standard NTSC, PAL and SECAM broadcast systems.
Line number range: 1~525 (NTSC) and 1~625 (PAL/SECAM)
Slope Trigger
Trigger condition
(, , =) Positive slope, (, , =) negative slope
Time setting
20ns~10s
Alternate Trigger
Trigger on CH1
Edge, Pulse, Video, Slope
Trigger on CH2
Edge, Pulse, Video, Slope
Pattern Trigger
[1]
Trigger mode
D0~D15 select H, L, X, ,
Duration Trigger
[1]
Trigger Type
D0~D15 select H, L, X
Qualifier
, , =
Time setup
20ns~10s
Measurements
Cursor
Manual
Voltage difference between cursors (V)
Time difference between cursors (T)
Reciprocal of T in Hertz (1/T)
Track
Voltage value for Y-axis waveform
Time value for X-axis waveform
Auto
Cursors are visible for Automatic Measurement
Auto Measure
Vpp, Vamp, Vmax, Vmin, Vtop, Vbase, Vavg, Vrms, Overshoot,
Preshoot, Freq, Period, Rise Time, Fall Time, +Width, -Width,
+Duty, -Duty, Delay12 , Delay12
Remarks:
[1] For DS1000D series;
[2] When sampling is 1GSa/s, only single channel can be used.