RIGOL Chapter 3 Performance Test
Service Guide for DS1000E, DS1000D Series
3-4
USB Device Interface Test
Purpose:
Test if the USB Device interface works normally through Ultrascope for DS1000E,
DS1000D series.
Tools:
A set of DS1000E, DS1000D series digital oscilloscope
A PC with USB interface
A standard USB cable (Type AB)
Ultrascope for DS1000E, DS1000D series
Steps:
1. Install the Ultrascope for DS1000E, DS1000D series software on the PC.
2. Connect the oscilloscope with PC using an USB cable, and install the driver
program step by step following the prompt.
Figure 3-3 USB Device interface connection
3. Run Ultrascope for DS1000E, DS1000D series, then click ToolsOptions and
select USB as the current IO.
Figure 3-4 Select USB interface