Rigol DM3058 Especificações Página 57

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 264
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 56
RIGOL
Programming Guide for DM3058
3-31
to Table 3-7.
Return Format:
The query returns the parameter related to current measuring rate: fast, medium
or slow.
Example:
Set the measuring rate of 2-wire resistance as Medium.
:rate:resistancemedium
6. :rate:fresistance
Command Format:
:rate:fresistance?
:rate:fresistance{fast|medium|slow}
Function:
The command queries or sets the measuring rate of 4-wire resistance.
Explanations:
:rate:fresistance?: Query current measuring rate of 4-wire resistance. 4WR
function has to be enabled before exacute this command.
:rate:fresistance{fast|medium|slow}: Set current measruing rate of 4-wire
resistance as Medium. For the range and explanation of parameters please refer
to Table 3-7.
Return Format:
The query returns the parameter related to current measuring rate: fast, medium
or slow.
Example:
Set the measruing rate of 4-wire resistance as Medium.
:rate:fresistancemedium