Chapter 6 Specifications RIGOL
User’s Guide for DM3068
6-5
Continuity/Diode Test
Response Time 300 samples/sec, with audible tone
Continuity Threshold Adjustable from 1 Ω to 2000 Ω
Auto Zero OFF Operation (typical value)
Following instrument warm-up at the environment temperature ±1 and <5 minutes, add 0.0001 % range + 2 uV error for DCV and 2 mΩ error for
resistance.
Settling Considerations
Reading settling times are affected by source impedance, cable dielectric characteristics and input signal changes. The default measurement delay can
ensure the correctness of the first reading for most measurements.
Measurement Considerations
Teflon or other high-impedance, low-dielectric absorption wire insulation is recommended for these measurements.