Programming Guide for DS1000B Series
2-35
3. :ACQuire:AVERages
Command Format:
:ACQuire:AVERages<count>
Function:
The command sets the average acquisition time. The <count> range is 2~256,
and the count increases by the multiples of 2.
Query Format:
:ACQuire:AVERages?
Return Format:
The query returns 2 or 4, 8, 16, 32, 64, 128, 256.
Example:
:ACQuire:AVERages16 Set the average acquisition time as 16.
:ACQuire:AVERages? Return 16.
4. :ACQuire:SRATe?
Command Format:
:ACQuire:SRATe? [{<CHANnel<n>}]
Function
To query sample rate of CHANnel <n>, <n> may be 1234.
Return Format:
The query returns 5.000e+05, the unit is Sa/s.