Rigol DS1204B Especificações Página 51

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 132
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 50
RIGOL
Programming Guide for DS1000B Series
2-43
1. :CHANnel<n>:BWLimit
Command Format:
:CHANnel<n>:BWLimit{{1|ON}|{0|OFF}}
Function:
The command sets bandwidth limit function ON (limit band width to 20MHz to
reduce noise) or OFF (full band width). The <n> may be 1, 2, 3 or 4.
Query Format:
:CHANnel<n>:BWLimit?
Return Format:
The query returns 1 or 0, respectively indicates ON or OFF.
Example:
:CHANnel2:BWLimiOFF Set BW limit function of channel 2 off.
:CHANnel2:BWLimit? Return 0.
2. :CHANnel<n>:COUPling
Command Format:
:CHANnel<n>:COUPling{DC | AC | GND}
Function:
The command sets coupling mode as DC (both AC and DC components of the
input signal can pass), AC (the DC component of the input signal can not pass)
or GND (disconnect the input signal). The <n> may be 1, 2, 3 or 4.
Query Format:
:CHANnel<n>:COUPling?
Return Format:
The query returns AC or DC, GND.
Example:
:CHANnel2:COUPlingDC Set the coupling mode of channel 2 as DC.
:CHANnel2:COUPling? Return DC.