2 Command System RIGOL
DS2000 Programming Guide
2-423
:TIMebase[:MAIN]:SCALe
Syntax
:TIMebase[:MAIN]:SCALe <scale_value>
:TIMebase[:MAIN]:SCALe?
Description
Set the scale of the main time base and the unit is s/div.
Query the current scale of the main time base.
Parameter
Name Type Range Default
<scale_value> Real
Depend on the time base
mode
[1]
:
Normal: 2ns
[2]
to 1ks
ROLL: 200ms to 1ks
1μs
Note
[1]
: refer to the :TIMebase:MODE command.
Note
[2]
: this value is different for different model. For DS2072 and DS2012, the value is 5 ns.
Return Format
The query returns the current scale of the main time base in scientific notation.
Example
:TIMebase:MAIN:SCALe 0.0002
The query returns 2.000000e-04.