Rigol DS1000B Manual do Utilizador Página 135

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 134
RIGOL
Users Guide for DS1000B Series
2-105
Table 2-88 The Cursor usage
Cursor Operation
Cursor A Turn the multifunctional knob ( ) to move cursor A horizontally.
Cursor B Turn the multifunctional knob ( ) to move Cursor B horizontally.
NOTE: Moving cursor horizontally is not allowed when other (not tracking cursor)
menu is activated.
4. To get measurement values:
Position of Cursor A: Time cursor centered on the midpoint of screen; Voltage
cursor centered on channel ground level.
Position of Cursor B: Time cursor centered on the midpoint of screen; Voltage
cursor centered on channel ground level.
Read the horizontal space between Cursor A and B (
X): Time between cursors,
units in seconds. (1/
X), units in Hz, kHz, MHz, GHz
Vertical space between cursor A and B (
Y): Voltage between cursors, units in V.
Vista de página 134
1 2 ... 130 131 132 133 134 135 136 137 138 139 140 ... 165 166

Comentários a estes Manuais

Sem comentários