Users Guide RIGOL
Publication number: UGC01135-1210
July 2010
DM3000 Series Digital Multimeter
DM3061/2/4
DM3051/2/4
© 2007 RIGOL Technologies, Inc. All Rights Reserved
Vista de página 0
1 2 3 4 5 6 ... 143 144

Resumo do Conteúdo

Página 1 - DM3051/2/4

User’s Guide RIGOL Publication number: UGC01135-1210 July 2010 DM3000 Series Digital Multimeter DM3061/2

Página 2

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series VIII Example 4: dB Measurement..

Página 3

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-52 Table 2- 39 Menu Description Function Menu Setting Description

Página 5

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-54 Delete or edit the setting task if needed. Figure 2- 58 View

Página 7 - General-Purpose Multimeter

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-56 Run the Scan Result Press Run in Main Interface of Scanning t

Página 9 - Content

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-58 1. Connect the test lead How to connect the lead in different

Página 11 - Chapter 1 Quick Start

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 3-4 Example 2: Elimination Leads Resistance Error When measuring

Página 14 - Appearance and Dimensions

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 3-6 Example 4: dB Measurement Do these steps as follows: 1. Connec

Página 16 - USB Device

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 3-8 Example 6: Temperature Sensor Setting a sensor is the same way o

Página 18

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 3-10 When you finish inputting all the data, press and save all th

Página 22

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 4-2 12. DCV is -1200V to 1200V In limit test and null value settings

Página 24

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-2 General Inspection Inspect a new DM3000 Digital Multimeter with

Página 25 - To Measure Resistance

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 4-4 To Change the Electric Power Fuse The electric power fuse locate

Página 27 - Terminal)

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-2 [1] Specifications are for 60 minutes warm–up and select 6½ reading r

Página 29 - To Measure Capacitance

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-4 [1] Specifications are for 60 minute warm–up and select 6½ reading res

Página 31 - To Test Continuity

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-6 Capacitance Characteristics Accuracy Specifications (% of reading + % of

Página 33 - To Check Diodes

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-8 Shunt Resistor: 0.025Ω for 1A, 10A, 1.025Ω for 200mA, 11.025Ω for

Página 36

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-10 Display: 256 x 64 pixels L

Página 38

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-12 [1] Specifications are for 60 minute warm–up, selecting 5 3/4 reading

Página 40

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-14 [1] Specifications are for 60 minute warm–up and select 5 3/4 reading

Página 42

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-16 Capacitance Characteristics Accuracy Specifications (%

Página 44 - To Choose Reading Resolution

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-18 Shunt Resistor: 0.025Ω for 1A, 10A, 1.025Ω for 200mA, 11.025Ω for

Página 46 - To Choose Range Options

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-4 Appearance and Dimensions Front Elevation Unit: mm Side

Página 47

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-20 General Specifications Display:

Página 49

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 6-2 Appendix B: Warranty RIGOL warrants that the products that it manufact

Página 51 - Continue Resistance

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 6-4 Appendix D: Contact RIGOL If you have any problem or requirement oc

Página 53 - DC Input Impedance

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-6 Figure 1- 4 The Rear Panel Figure 1- 5 Interface Exp

Página 55

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-8 test leads into circuit and start to measure. Figure 1- 7 DC Vo

Página 58 - Frequency Measurement

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-10 (To know the specific setting methods of the Null valu

Página 60 - Math Functions

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-12 Figure 1- 13 DC Current Measurement 6. Use history function.

Página 62 - Statistic Measurement

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-14 user demand. If user does not implement Null computing, this pa

Página 64

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-16 Null computing will be an option operation, could be setup in a

Página 66

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-18 Figure 1- 22 4-Wire Resistance Measurement 6. Use history f

Página 69

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-20 Figure 1- 25 Capacitance Measurement 6. Use history function

Página 71 - External Triggering

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-22 Figure 1- 28 Continuity Measurement I Open or Closed Circuit

Página 73 - Store and Recall

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-24 3. Lead test leads into circuit and start to check. Figure

Página 75 - File Operation

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-26 4. Lead test leads into circuit and start to check. Figure 1-

Página 77 - Set Up the I/O System

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-28 4. Lead test leads into circuit, start to check. Figure 1- 35

Página 79 - Manual IP Setting Interface

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series II Safety Notices Review the fo

Página 80

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-30 Figure 1- 39 Set Property of the Sensor  Press Name to crea

Página 82 - System settings

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-32 Figure 1- 48 Save or Apply Interface  Press Save to save th

Página 84 - Set Up the Display

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-34 To Choose Reading Resolution The measurement reading resolutio

Página 86 - Set Up System date/time

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-36 To Choose Range Options Use “manual” selection or “automat

Página 88 - Factory Defaults Setting

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-38 To Control Trigger Options Use or to trigger the multime

Página 91

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-2 To Set up Measurement Parameters Press key to enter the

Página 93 - 1. Data Log rate

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-4 AC Filter There are three settings at the AC Filter menu. Choose

Página 95

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-6 Null Measurement The DM3000 null settings is available for DC vo

Página 97

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-8 Ratio Measurement Use Ratio measurement to measure the rat

Página 99

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-10 Frequency Measurement Frequency measurement function is used fo

Página 101

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series IV Safety Terms and Symbols Term

Página 102

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-12 Math Functions The DM3000 provides five math functions: Null,

Página 104 - View the Scan Result

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-14 Statistic Measurement The Statistic function is for DC voltage,

Página 106

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-16 dB Measurement The dB function applies to AC voltage and DC vol

Página 108

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-18 To Set up Triggering Parameter Function The DM3000 triggering

Página 110 - Example 4: dB Measurement

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-20 Reading Hold The reading hold mode captures and holds a readin

Página 113

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-22 Single Triggering The multimeter takes one reading, or a number

Página 115 - Example 7: Reading Hold

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-24 To Set up the VMC At external triggering mode, when the data sa

Página 117 - Prompting Message

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-26 File Storage In local/U-disk storage area, you allowed to save,

Página 119 - Troubleshooting

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-28 To Set up the Utility The Utility function establishes sys

Página 121 - Chapter 5 Characteristics

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-30 1. To Setup the LAN Parameters Set up a LAN configuration and r

Página 123 - AC Characteristics

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series VI With the performance and cha

Página 124

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-32 2. To Set Up the GPIB Parameter Each device on the GPIB (IEE

Página 126 - Capacitance Characteristics

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-34 System settings Press  System, enter the menu shown below:

Página 128

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-36 Set Up the Display Press  Sys  Disp, enter the menu shown

Página 130

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-38 Set Up System date/time Press  Sys  Clock, enter the menu

Página 132

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-40 Factory Defaults Setting Press  System  Cfg  Default, the

Página 135 - 28 ℃ to 55 ℃

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-42 High-Speed Data Log High-speed data log features: display mode

Página 137

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-44 Figure 2- 44 Start Delay Setting Press  Log  Stop, the

Página 139

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-46 2. The Data Log Trigger and Delay To set the start condition

Página 141 - Chapter 6 Appendix

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-48 3. The Data log stop To set the conditions to stop Data Lo

Página 143

RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-50 Multi-Route Scanning DM3054 and DM3064 have a built-in inspect