User’s Guide for DM3000 Series
2-46
2. The Data Log Trigger and Delay
To set the start condition and the delay time of Data Log.
Press Log Start, the interface shows below.
Figure 2- 47 The Menu of Start Manner Interface
Table 2- 36 Menu Description
Function
Menu
Setting
Description
Trig
Manu
Ext
Set trigger: Manual
Set trigger: External
Delay
The time from the start of trigger to the
commence of data logging.
Save all the changes, back to a higher
level menu.
External trigger
The multimeter receives a trigger signal from the “Ext Trig” at the rear panel.
Once at External trigger mode the other trigger is prohibited.
Manual trigger
Press under the Manual trigger mode to obtain continuous data.
The factory default trigger mode is Manual trigger.
Start delay
The start delay time refers to the time from the first sample to the start of the second
sample.
Press Log Start Delay, the interface shows below.