Rigol M300 Manual do Utilizador Página 19

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 18
RIGOL
© 2007 RIGOL Technologies, Inc.
User’s Guide for DM3000 Series
1-9
To Measure AC Voltage
The following shows the system connections and selection of measurement functions.
This practice provides a guide to get familiar with the AC Voltage measurement
technique. (The AC functions only support 5½ digits measurement.)
Figure 1- 9 AC Voltage Measurement Data Interface
Table1- 2 AC Voltage Measurement Characteristics
Five Ranges
200mV, 2V, 20V, 200V, 750V
Max Resolution
100nV
Input Protection
750VRMS on all ranges (HI Terminal)
Configurable
Parameters
Range, AC Filter, Null value
Steps:
1. Connect test leads as shown in Figure 1- 10; red test lead to the HI Terminal,
black test lead to the LO Terminal.
2. Press to select the AC Voltage measurement function.
3. Choose the appropriate measurement range.
4. Setup the AC Filter.
Press Filter, to setup the AC Filter Bandwidth (Default value: Mid).
5. Set the Null value.
Null computing will be an option operation, could be setup in accordance with
user demand. If user does not implement Null computing, this parameter is not
required, direct implementation of the next step.
Vista de página 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 143 144

Comentários a estes Manuais

Sem comentários