Rigol DM3058 Especificações Página 145

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 264
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 144
RIGOL
Programming Guide for DM3058
5-29
Return Format:
The query returns the data in ouput buffer.
3. MEASure:CURRent:AC?
Command Format:
MEASure:CURRent:AC?
[{<range>|AUTO|MIN|MAX|DEF} [,{<resolution>|MIN|MAX|DEF}]]
Function:
The command resets all the AC current measuring parameters and trigger
parameters as defaults, and sets AC current measuring function, get the
measuring results and send them to output buffer of instrument.
Parameters:
Name
Type
Range of Values
Default
<range>
Discrete
{ 20 mA | 200 mA
| 2 A | 10 A }
200 mA
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
<range> may be 20 mA (MIN), 200 mA, 1 A, 10 A (MAX), AUTO (autorange) (unit
is A). MAX = 10, MIN = 0. 02.
<resolution> (unit is A); in default setting, the rate is medium.
Return Format:
The query returns the data in output buffer.
4. MEASure:CURRent[:DC]?
Command Format:
MEASure:CURRent[:DC]?
[{<range>|AUTO|MIN|MAX|DEF} [,{<resolution>|MIN|MAX|DEF}]]
Function:
The command resets all the DC current measuring parameters and trigger
parameters as defaults, and sets DC current measuring function, gets the
Vista de página 144
1 2 ... 140 141 142 143 144 145 146 147 148 149 150 ... 263 264

Comentários a estes Manuais

Sem comentários