Explanations:
[SENSe:]FRESistance:RESolution {<resolution>|MIN|MAX|DEF}: Set the
resolution of 4WR measurement. In default of <resolution>, the rate is medium.
[SENSe:]FRESistance:RESolution?[{MIN|MAX}: Query the resolution of 4WR
measurement.
2-wire and 4-wire resistance function both are influenced by the command. All
the command of 2WR and 4WR will have influence over the setting of resistance.
Return Format:
The query returns the resolution of 4WR measurement.
52. [SENSe:]FREQuency:APERture
Command Format:
[SENSe:]FREQuency:APERture{<seconds>|MIN|MAX|DEF}
[SENSe:]FREQuency:APERture?[{MIN|MAX}]
Function:
The command sets or queries the interval time of frequency measurement.
Parameters:
Name
Type
Range of Values
Default
<seconds>
Consecutive
Real Number
400 ms~2 s
400 ms
Explanations:
[SENSe:]FREQuency:APERture{<seconds>|MIN|MAX|DEF}: Set the interval
time of frequency measurement. Unit is s, so 400 ms should be input in form of
0.4.
[SENSe:]FREQuency:APERture? [{MIN|MAX}]: Query the interval time of
frequency measurement.
Return Format:
The query returns the interval time of frequency measurement.
53. [SENSe:]FREQuency:NULL[:STATe]