RIGOL
Programming Guide for DM3058
5-33
Function:
The command resets all the 2-wire resistance measuring parameters and trigger
parameters as defaults, and sets 2-wire resistance measuring function, get the
measuring results and send them to output buffer of instrument.
Parameters:
Name
Type
Range of Values
Default
<range>
Discrete
{200 Ω | 2 kΩ | 20 kΩ | 200
| 2 MΩ | 10 MΩ | 100 MΩ}
200
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
<range> may be 200 Ω (MIN), 2 , 20 , 200 , 2 , 10 , 100 (MAX),
AUTO (autorange) (unit is Ω). MAX = 100 000 000, MIN = 200.
In default setting of <resolution>, the rate is medium.
Return Format:
The query returns the data in output buffer.
10. MEASure[:VOLTage]:AC?
Command Format:
MEASure[:VOLTage]:AC?
[{<range>|AUTO|MIN|MAX|DEF} [,{<resolution>|MIN|MAX|DEF}]]
Function:
The command resets all the AC voltage measuring parameters and trigger
parameters as defaults, and sets AC voltage measuring function, gets the
measuring results and sends them to output buffer of instrument.
Parameters:
Name
Type
Range of Values
Default
<range>
Discrete
{ 200 mV | 2 V | 20 V
| 200 V | 750 V }
20 V
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
Vista de página 148