RIGOL
Programming Guide for DM3058
5-67
measurement.
Return Format:
The query returns the resolution of DCV measurement.
32. [SENSe:]VOLTage[:DC]:ZERO:AUTO
Command Format:
[SENSe:]VOLTage[:DC]:ZERO:AUTO{ON|OFF}
[SENSe:]VOLTage[:DC]:ZERO:AUTO?
Function:
The command sets or queries the enable state of DCV auto zero measuring
function.
Explanations:
[SENSe:]VOLTage[:DC]:ZERO:AUTO{ON|OFF}: Enable or disable DCV auto
zero measuring function.
[SENSe:]VOLTage[:DC]:ZERO:AUTO?: Query the enable state of DCV auto zero
measuring function.
Return Format:
The query returns the enable state of DCV auto zero measuring function.
33. [SENSe:]RESistance:APERture
Command Format:
[SENSe:]RESistance:APERture{<seconds>|MIN|MAX|DEF}
[SENSe:]RESistance:APERture?[{MIN|MAX}]
Function:
The command sets or queries the interval time of 2-wire resistance
measurement.