2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-97
:DISPlay:GRADing:TIME
Syntax
:DISPlay:GRADing:TIME <time>
:DISPlay:GRADing:TIME?
Description
Set or query the persistence time and the unit is s.
Parameter
Name Type Range Default
<time> Discrete {MIN|0.05|0.1|0.2|0.5|1|2|5|10|20|INFinite} MIN
Explanation
MIN: set the persistence time to its minimum to view the waveform changing in high refresh rate.
Specific Values: a certain value between 0.05s and 20s, enable to observe glitch that changes
relatively slowly or glitch with low occurrence probability.
INFinite: in this mode, the oscilloscope displays the newly acquired waveform without clearing the
waveform formerly acquired. Enable to measure noise and jitter as well as capture incidental events.
Return Format
The query returns the persistence time set.
:DISPlay:WBRightness
Syntax
:DISPlay:WBRightness <time>
:DISPlay:WBRightness?
Description
Set or query the waveform brightness and the unit is %.
Parameter
Name Type Range Default
<time> Integer 0 to 100 50
Return Format
The query returns an integer between 0 and 100.