Rigol DS2000A Series Especificações Página 320

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 390
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 319
RIGOL 2 Command System
2-304 MSO2000A/DS2000A Programming Guide
:TRIGger:PULSe
:TRIGger:PULSe:LEVel
:TRIGger:PULSe:LWIDth
:TRIGger:PULSe:SOURce
:TRIGger:PULSe:UWIDth
:TRIGger:PULSe:WHEN
:TRIGger:PULSe:LEVel
Syntax
:TRIGger:PULSe:LEVel <level>
:TRIGger:PULSe:LEVel?
Description
Set or query the trigger level in pulse trigger when the trigger source is CHANnel1, CHANnel2 or EXT. The
unit is the same with the current amplitude unit.
Parameter
Name Type Range Default
<level> Real
When the trigger source is CHANnel1 or CHANnel2:
-5×VerticalScaleOFFSet to +5×VerticalScale–OFFSet
When the trigger source is EXT: -4V to 4V
0
Note: For the VerticalScale, refer to the :CHANnel<n>:SCALe command. For the OFFSet, refer to
the
:CHANnel<n>:OFFSet command.
Explanation
To set the trigger source of pulse trigger, please refer to the
:TRIGger:PULSe:SOURce command.
When the trigger source of pulse trigger is a digital channel, please use the
:LA:POD<n>:THReshold
command to set or query the threshold level of pulse trigger.
Return Format
The query returns the trigger level in pulse trigger when the trigger source is CHANnel1, CHANnel2 or EXT
in scientific notation.
Vista de página 319
1 2 ... 315 316 317 318 319 320 321 322 323 324 325 ... 389 390

Comentários a estes Manuais