Rigol DS2000A Series Especificações Página 361

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 390
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 360
2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-345
:WAVeform:MODE
Syntax
:WAVeform:MODE <mode>
:WAVeform:MODE?
Description
Set or query the data reading mode of the
:WAVeform:DATA? Command.
Parameter
Name Type Range Default
<mode> Discrete {NORMal|RAW|MAXimum} NORMal
Explanation
NORMal: read the waveform data currently displayed on the screen.
RAW: read the waveform data in the internal memory.
Note: The waveform data in the internal memory can only be read when the oscilloscope is in the
stop state and please do not operate the oscilloscope during the reading process.
MAXimum: in the run state, read the waveform data displayed on the screen; in the stop state, read
the waveform data in the internal memory.
Return Format
The query returns NORM, MAX or RAW.
Vista de página 360
1 2 ... 356 357 358 359 360 361 362 363 364 365 366 ... 389 390

Comentários a estes Manuais

Sem comentários