RIGOL Chapter 2 Command System
2-134 DS1000Z Programming Guide
:WAVeform:MODE
Syntax :WAVeform:MODE <mode>
:WAVeform:MODE?
Description
Set or query the reading mode used by :WAVeform:DATA?.
Parameter
Name
Type
Range
Default
<mode> Discrete {NORMal|MAXimum|RAW} NORMal
Explanation NORMal: read the waveform data displayed on the screen.
MAXimum: read the waveform data displayed on the screen when the instrument is
in run state and the waveform data in the internal memory in stop state.
RAW: read the waveform data in the internal memory. Note that the waveform data
in the internal memory can only be read when the oscilloscope is in stop state and
the oscilloscope can not be operated.
If the MATH channel is selected, only NORMal mode is valid.
Return
Format
The query returns NORM, MAX or RAW.
Example
:WAVeform:MODE RAW /*Set the reading mode to RAW*/
Related
Command
:WAVeform:DATA?
:WAVeform:FORMat
Syntax :WAVeform:FORMat <format>
:WAVeform:FORMat?
Description
Set or query the return format of the waveform data.
Parameter
Name
Type
Range
Default
<format> Discrete {WORD|BYTE|ASCii} BYTE
Explanation BYTE: a waveform point occupies one byte (namely 8 bits).
WORD: a waveform point occupies two bytes (namely 16 bits) in which the lower 8
bits is valid and the higher 8 bits are 0.
ASCii:
return the waveform points in character number. Waveform points are retuned
in scientific notation and separated by comma.
Return
Format
The query returns WORD, BYTE or ASC.
Example
:WAVeform:FORMat WORD /*Set the return format to WORD*/