Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-47
:MASK:MDISplay
Syntax :MASK:MDISplay <bool>
:MASK:MDISplay?
Description Enable or disable the statistic information when the pass/fail test is enabled, or query the
status of the statistic information.
Parameter
Name
Type
Range
Default
<bool>
Bool
{{0|OFF}|{1|ON}}
0|OFF
Explanation Before executing this command, you need to sent the :MASK:ENABle command to
enable the pass/fail test.
When the statistic information is on, the test results will be displayed at the upper
right corner of the screen.
You can sent
:MASK:PASSed?, :MASK:FAILed? and :MASK:TOTal? to query the test
results.
Return
Format
The query returns 0 or 1.
Example
:MASK:MDISplay ON
Related
Command
:MASK:ENABle
:MASK:PASSed?
:MASK:FAILed?
:MASK:TOTal?
:MASK:SOOutput
Syntax :MASK:SOOutput <bool>
:MASK:SOOutput?
Description Turn the “Stop on Failfunction on or off, or query the status of the “Stop on Fail
function.
Parameter
Name
Type
Range
Default
<bool>
Bool
{{0|OFF}|{1|ON}}
0|OFF
Explanation ON: when failed waveforms are detected, the oscilloscope will stop the test and enter
the STOPstate. At this point, the results of the test remain the same on the screen
(if display is turned on) and only one pulse is output from the [Trigger Out]
connector (if enabled) at the rear panel.
OFF: the oscilloscope will continue with the test even though failed waveforms are
detected. The test results on the screen will update continuously and a pulse will be
output from the [Trigger Out] connector at the rear panel each time a failed
waveform is detected.
Return
Format
The query returns 0 or 1.
Example
:MASK:SOOutput ON