RIGOL
Programming Guide for DM3058/DM3058E
4-31
1. [SENSe:]FUNCtion
Command Format:
[SENSe:]FUNCtion "<function>"
[SENSe:]FUNCtion?
Function:
The commands set and query the current measurement function.
Parameter:
Name Type Range Default
<function> Discrete Refer to Table 4-3
—
Explanations:
[SENSe:]FUNCtion "<function>" – Selects the measurement function to be used.
Note this comand has no default value and <function> must be set, see table
below.
[SENSe:]FUNCtion? – Queries the current measurement function.
Table 4-3 Parameters and explanations
Parameter Meaning Parameter Meaning
VOLTage:DC DC voltage PERiod Period
VOLTage:AC AC voltage RESistance 2WR
CURRent:AC AC current FRESistance 4WR
CURRent:DC DC current CONTinuity Continuity
FREQuency Frequency DIODe
Diode
Return Format:
The query returns the current measurement function.
2. [SENSe:]VOLTage:DC:RANGe
Command Format:
[SENSe:]VOLTage:DC:RANGe {<range>|MINimum|MAXimum}
[SENSe:]VOLTage:DC:RANGe? [MINimum|MAXimum]
Function: