Rigol DM3058 Manual do Utilizador Página 137

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 192
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 136
RIGOL
Programming Guide for DM3058/DM3058E
4-44
Parameter:
Name Type Range Default
<resolution> Discrete Refer to Table 4-2 1ppm x range
Explanations:
[SENSe:]CURRent:AC:RESolution {<resolution>|MINimum|MAXimum} Sets
the desired precision of AC current measurement
[SENSe:]CURRent:AC:RESolution? [MINimum|MAXimum] Queries the current
precision of AC current measurement.
<resolution> relates to measurement rate and integration time, refer to Table
4-2.
Return Format:
The query returns the current precision of AC current measurement.
22. [SENSe:]RESistance:RESolution
Command Format:
[SENSe:]RESistance:RESolution {<resolution>|MINimum|MAXimum}
[SENSe:]RESistance:RESolution? [MINimum|MAXimum]
Function:
The commands set and query the precision of 2-wire resistance measurement.
Parameter:
Name Type Range Default
<resolution> Discrete Refer to Table 4-2 3ppm x range
Explanations:
[SENSe:]RESistance:RESolution {<resolution>|MINimum|MAXimum} Sets the
desired precision of 2-wire resistance measurement
[SENSe:]RESistance:RESolution? [MINimum|MAXimum] Queries the current
precision of 2-wire resistance measurement.
<resolution> relates to measurement rate and integration time, refer to Table
4-2.
Both 2-wire and 4-wire resistance measurements are influenced by this command.
All the commands of 2-wire and 4-wire resistance will have influence over the
Vista de página 136
1 2 ... 132 133 134 135 136 137 138 139 140 141 142 ... 191 192

Comentários a estes Manuais

Sem comentários