Explanations:
When the measurement result is within the maximum and minimum, the PF
operation is success.
Return Format:
The query returns the PF operation: PASS, HI or LO.
16. :CALCulate:PF:LOWEr
Command Format:
:CALCulate:PF:LOWEr?
:CALCulate:PF:LOWEr {<range>|MIN|MAX|DEF}
Function:
The comands query and set the lower limit of P/F operation.
Parameter:
Name Type Range Default
<range>
Consecutive
real numbe
See Table 3-3 0
Explanations:
:CALCulate:PF:LOWEr?Queries the lower limit of P/F operation
:CALCulate:PF:LOWEr {<range>|MIN|MAX|DEF} Sets the lower limit of P/F
operation. See table below for more details.
Table 3-3 Ranges of P/F operation
Measurent
items
Range Unit
DC Voltage
±1200 V
AC Voltage 0~900 V
DC Current ±12 A
AC Current 0~12 A
Resistance
0~1.2e+08 Ω
Capacitance 0~1.2e-02 F
Period 1.0e-06~100 s
Frequency 0~1.2e+06 Hz