Programming Guide for DM3058/DM3058E
6-18
25 Return 4
26 *cls
27 :function:diode
28 :calculate:statistic:min? /*Beeping*/
29 SYST:ERR?
30 Return setting unacceptable
31 *esr?
32 Return 8
Explanations:
1. Commands in line 2, 9, 15, 21 and 26 clear the current status of the status
register, the value of the event status register and the error queue. Execute “*cls
every time when you input new command for quering the register status, in order
to suppress interference by information existed in register.
2. Commands in line 4, 10, 16, 22, 28 wigh beeping indicate the wrong operation.
3. SYST:ERR? is used to query the current error information.
4. *esr? is for query the error type, refer to Figure 2-1.
5. For line 4, the data returnedl covers the data in the last time and generate a
query interrupted” error. It is a query error.
6. For line 10, the character “*”
attached before the command may cause a “synax
error” error. It is a command error.
7. For line 16, the parameters should only be {RIGOL|AGILENT|FLUKE}, so null
parameter is invalid, a “Parameter error” is generated. It is an execution error.
8. For line 22, the command has no data returned. If read the data forcibly, a “Query
Unterminated” error will generate. It is a query error.
9. The command in line 28 queries the calculated minimum value, it should not be
used in continuity test and diode measurement. Therefore, this command is
wrong and which may disable the instrument and generate a “setting
unacceptable” error. It is a device error.