RIGOL
Programming Guide for DM3058/DM3058E
4-39
13. [SENSe:]CURRent:AC:RANGe:AUTO
Command Format:
[SENSe:]CURRent:AC:RANGe:AUTO {ON|OFF}
[SENSe:]CURRent:AC:RANGe:AUTO?
Function:
The commands set and query the auto range state of AC current measurement.
Explanations:
[SENSe:]CURRent:AC:RANGe:AUTO {ON|OFF}} - Sets the desired auto range
state. If choose ON, the instrument will select a measurement range according to
the signal inputted.
[SENSe:]CURRent:AC:RANGe:AUTO? Queries the current auto range state of
AC current measurement.
Return Format:
The query returns the auto range state of AC current measurement.
14. [SENSe:]RESistance:RANGe:AUTO
Command Format:
[SENSe:]RESistance:RANGe:AUTO {ON|OFF}
[SENSe:]RESistance:RANGe:AUTO?
Function:
The commands set and query the auto range state of 2-wire resistance
measurement.
Explanations:
[SENSe:]RESistance:RANGe:AUTO {ON|OFF} - Sets the desired auto range state
of 2-wire resistance measurement. If choose ON, the instrument will select a
measurement range according to the signal inputted.
[SENSe:]RESistance:RANGe:AUTO? Queries the current auto range state of
2-wire resistance measurement.
Both 2-wire and 4-wire resistance measurements are influenced by this command.
All the commands of 2-wire and 4-wire resistance will have influence over the
setting of resistance.