RIGOL Chapter 2 Command System
Programming Guide for DM3000
Syntax
:MEASure:PERiod?
Function The query returns the period measurement value in scientific notation
(for example, +2.77679688E-03) and the unit is s.
44.
:MEASure:PERiod
Syntax
:MEASure:PERiod {0|1|2|3|4|MIN|MAX|DEF}
Function Set the period measurement range.
Explanation For the meaning of each range label, please refer to the
“Explanation” in :MEASure:VOLTage:AC.
The period measurement ranges from 3.3 us to 0.33 s.
Example Set the period measurement range to the maximum:
:MEASure:PERiod 4 or
:MEASure:PERiod MAX
45.
:MEASure:PERiod:RANGe?
Syntax
:MEASure:PERiod:RANGe?
Function Query the AC voltage range currently used by period measurement.
The query returns 0, 1, 2, 3 or 4.
46. :MEASure:PERiod:DIGIt
Syntax :MEASure:PERiod:DIGIt?
:MEASure:PERiod:DIGIt {INC|DEC|5|6|7}
Function Set the display digit of period measurement value.
The query returns 5, 6 or 7.
Explanation
Refer to the “Explanation” in :MEASure:VOLTage:DC:DIGIt.
Example Set the display digit of period measurement value to 7:
:MEASure:PERiod:DIGIt 7
Decrease the display digit by one:
:MEASure:PERiod:DIGIt DEC
47.
:MEASure:CONTinuity?
Syntax
:MEASure:CONTinuity?
Function The query returns the resistance connected to the multimeter under
continuity measurement in scientific notation and the unit is Ω.
48.
:MEASure:CONTinuity