Rigol M300 Manual do Utilizador Página 78

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 89
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 77
RIGOL Chapter 3 Commands Compatibility
Programming Guide for DM3000
3-12
{<range>|MIN|MAX|DEF}
[SENSe:]RESistance:RESolution?
[{MIN|MAX}]
[SENSe:]RESistance:RESolution
{<resolution>|MIN|MAX|DEF}
Query and set the reading resolution of 2-wire
resistance measurement.
[SENSe:]VOLTage:AC:BANDwidth?
[{MIN|MAX}]
[SENSe:]VOLTage:AC:BANDwidth
{<filter>|MIN|MAX|DEF}
Query the filter bandwidth of AC voltage
measurement and the query returns the boundary
bandwidth (3, 20 or 200).
Set the filter bandwidth of AC voltage measurement
to 3, 20 or 200.
[SENSe:]VOLTage:AC:NULL[:STATe]?
[SENSe:]VOLTage:AC:NULL[:STATe]
{ON|OFF}
Query and set the state of the NULL function of AC
voltage measurement.
[SENSe:]VOLTage:AC:NULL:VALue?
[{MIN|MAX}]
[SENSe:]VOLTage:AC:NULL:VALue
{<value>|MIN|MAX}
Query and set the NULL value of AC voltage
measurement.
[SENSe:]VOLTage:AC:PEAK:STATe?
[SENSe:]VOLTage:AC:PEAK:STATe {ON|OFF}
Query and set the state of the peak-peak
measurement function of AC voltage measurement.
Note: DM3000 only receives the command but
does not support the peak-peak measurement
function.
[SENSe:]VOLTage:AC:RANGe:AUTO?
[SENSe:]VOLTage:AC:RANGe:AUTO <mode>
Query and set the state of the auto range function
of AC voltage measurement.
[SENSe:]VOLTage:AC:RANGe[:UPPer]?
[{MIN|MAX}]
[SENSe:]VOLTage:AC:RANGe[:UPPer]
{<range>|MIN|MAX|DEF}
Query and set the range of AC voltage
measurement.
[SENSe:]VOLTage[:DC]:APERture?
[{MIN|MAX}]
[SENSe:]VOLTage[:DC]:APERture
{<second>|MIN|MAX|DEF}
Query and set the aperture time of DC voltage
measurement.
Note: DM3000 only receives the command but
does not respond to it.
[SENSe:]VOLTage[:DC]:APERture:ENABle?
[SENSe:]VOLTage[:DC]:APERture:ENABle
{ON}
Query the state of the interval function of DC
voltage measurement.
Set the interval function of DC voltage
measurement toON.
[SENSe:]VOLTage[:DC]:IMPedance:AUTO? DM3000 only receives this command but does not