Rigol M300 Manual do Utilizador Página 49

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 89
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 48
Chapter 2 Command System RIGOL
Programming Guide for DM3000
2-37
:TRIGger:AUTO:HOLD {ON|OFF|1|0}
Function Enable or disable the auto trigger delay function.
The query returns ON or OFF.
Default
OFF
4.
:TRIGger:AUTO:HOLD:SENSitivity
Syntax :TRIGger:AUTO:HOLD:SENSitivity?
:TRIGger:AUTO:HOLD:SENSitivity {0|1|2|3|MIN|MAX|DEF}
Function Set the sensitivity of the auto trigger delay.
The query returns 0, 1, 2 or 3.
Explanation The sensitivity corresponding to each parameter is as follows.
Value Sensitivity
0 0.01%
1 0.1%
2 1%
3 10%
MAX 10%
MIN 0.01%
DEF 0.1%
5.
:TRIGger:SINGle
Syntax :TRIGger:SINGle?
:TRIGger:SINGle {
<value>
|MIN|MAX|DEF}
Function
Set the number of samples for single trigger.
Explanation
<value>
ranges from 1 to 1000.
The DEFis 1.
6. :TRIGger:SINGle:TRIGger
Syntax
:TRIGger:SINGle:TRIGger
Function Execute single trigger.
Explanation First, select the single trigger source using the :TRIGger:SOURce
SINGLE command; otherwise, this command is invalid.
7.
:TRIGger:EXT
Syntax :TRIGger:EXT?
:TRIGger:EXT {RISE|FALL}
Function
Set the external trigger type to RISE or FALL.
Vista de página 48