RIGOL Chapter 2 Command System
Programming Guide for DM3000
1. :UTILity:INTErface:LAN:DHCP
Syntax :UTILity:INTErface:LAN:DHCP?
:UTILity:INTErface:LAN:DHCP {ON|OFF|1|0}
Function Enable or disable the DHCP setting.
The query returns ON or OFF.
Explanation The three modes (DHCP, Auto IP and Manual IP) are not permitted
be disabled at the same time.
Default ON
2.
:UTILity:INTErface:LAN:AUTOip
Syntax :UTILity:INTErface:LAN:AUTOip?
:UTILity:INTErface:LAN:AUTOip {ON|OFF|1|0}
Function Enable or disable the AutoIP setting.
The query returns ON or OFF.
Explanation
The three modes (DHCP, Auto IP and Manual IP) are not permitted
be disabled at the same time.
Default
ON
3. :UTILity:INTErface:LAN:MANUip
Syntax :UTILity:INTErface:LAN:MANUip?
:UTILity:INTErface:LAN:MANUip {ON|OFF|1|0}
Function
Enable or disable the ManualIP setting.
Explanation
The three modes (DHCP, Auto IP and Manual IP) are not permitted
be disabled at the same time.
Default OFF
4.
:UTILity:INTErface:LAN:IP
Syntax :UTILity:INTErface:LAN:IP?
:UTILity:INTErface:LAN:IP
<ip_address>
Function Users define the IP address of the multimeter.
Explanation The format of
<ip_address>
is “nnn.nnn.nnn.nnn”. The first
“nnn”
ranges from 0 to 223 (except 127) and the others range
from 0 to 255.
The IP address configuration type should be Manual and both
DHCP and AutoIP should be disabled when you use this
command.
Comentários a estes Manuais